Chimney Stack Directivity Correction

Chimney Stack Directivity Correction

The calculation below is taken from Annex B — Table B.1 of ISO 9613-2:2024. The calculation provides the directivity correction ($D_c$) for chimney stacks as a function of two values:

  • Radiation angle ($\theta$) relative to the vertical stack axis
  • The Parameter ($ka$) (dimensionless product of wave number and chimney radius)

$D_c$ (dB) is the angle in degrees between vertical and the receiver.

The parameter $ka$ is calculated from:

$$ ka = \frac{2 \pi a f}{331.4 \sqrt{1 + \frac{T}{273}}} $$

Where:

  • $T$: Temperature at the chimney mouth
  • $f$: Frequency in Hz
  • $a$: Opening radius
  • $c$: Speed of sound (343 ms-1 typically)

So, before you use the table below you will have:

  • One angle value; and
  • Several ka values (one for each frequency considered).

There will be a $ka$ value for each frequency band you are calculating for.

Use the lookup table below to calculate the attenuation in dB for each frequency ($ka$ and $D_c$ pair)

θ (deg) $ka$=4.0 $ka$=5.0 $ka$=6.3 $ka$=8.0 $ka$=10.1 $ka$=12.7 $ka$=16.0 $ka$=20.2 $ka$=25.4 $ka$=32.0
30 2.4 2.1 1.9 2.0 2.1 2.6 3.1 3.4 3.4 3.3
45 4.0 3.4 3.1 3.1 3.4 4.0 4.4 4.6 4.6 4.5
60 4.0 3.4 3.1 3.1 3.4 4.0 4.4 4.6 4.6 4.5
75 2.4 2.1 1.9 2.0 2.1 2.6 3.1 3.4 3.4 3.3
90 -2.4 -2.2 -2.0 -1.9 -1.9 -1.9 -1.9 -2.1 -2.3 -2.7
105 -4.3 -4.6 -5.0 -5.4 -5.9 -6.4 -6.9 -7.3 -7.6 -7.9
120 -6.3 -7.0 -7.7 -8.2 -8.7 -9.1 -9.6 -10.2 -11.0 -12.1
  • For $ka > 32$ just use the ‘32’ column
  • For $\theta < 30^\circ$ use the ‘30’ row.
  • For $\theta > 120^\circ$ use the ‘120’ row.

Note that:

  • The data are based on measurements of large brick industrial chimneys.
  • They represent downwind propagation conditions (~3 m/s).
  • For chimneys of different design the results should be treated as approximate.